tech / sci.electronics.design / Low level AC measuremets with HP 34401

SubjectAuthor
* Low level AC measuremets with HP 34401Reinhard Zwirner
+* Re: Low level AC measuremets with HP 34401Lasse Langwadt Christensen
|`- Re: Low level AC measuremets with HP 34401Don
+* Re: Low level AC measuremets with HP 34401jlarkin
|`* Re: Low level AC measuremets with HP 34401Klaus Kragelund
| `* Re: Low level AC measuremets with HP 34401Cydrome Leader
|  +* Re: Low level AC measuremets with HP 34401Klaus Kragelund
|  |`- Re: Low level AC measuremets with HP 34401Cydrome Leader
|  `- Re: Low level AC measuremets with HP 34401John Larkin
+* Re: Low level AC measuremets with HP 34401Phil Allison
|`- Re: Low level AC measuremets with HP 34401Phil Allison
+- Re: Low level AC measuremets with HP 34401Don Y
`- Thank you! (was: Low level AC measuremets with HP 34401)Reinhard Zwirner

1
Low level AC measuremets with HP 34401

<j2i36iFhbirU1@mid.individual.net>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85589&group=sci.electronics.design#85589

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!weretis.net!feeder8.news.weretis.net!news.mixmin.net!news2.arglkargh.de!news.karotte.org!fu-berlin.de!uni-berlin.de!individual.net!not-for-mail
From: reinhard...@t-online.de (Reinhard Zwirner)
Newsgroups: sci.electronics.design
Subject: Low level AC measuremets with HP 34401
Date: Thu, 23 Dec 2021 02:54:27 +0100
Lines: 16
Message-ID: <j2i36iFhbirU1@mid.individual.net>
Mime-Version: 1.0
Content-Type: text/plain; charset=UTF-8
Content-Transfer-Encoding: 7bit
X-Trace: individual.net k3+C5dFwxFTzU0SBU6pfcwwFQFgcvCZGkeF54zo1ZhJ2Q=
Cancel-Lock: sha1:9cJxmM9EBF/LvwTPHhq9IrfIEZY=
X-Mozilla-News-Host: news://news.individual.de:119
User-Agent: Mozilla/5.0 (Windows NT 10.0; Win64; x64; rv:68.0) Gecko/20100101
Firefox/68.0 SeaMonkey/2.53.10.1
 by: Reinhard Zwirner - Thu, 23 Dec 2021 01:54 UTC

Hello,

Due to the used RMS converter low level AC measurememts with this DMM
are less precise to incorrect. There is a lab note as to this problem
by Gellerlabs containing three pictures. I have this note which,
unfortunately, lacks the mentioned pictures. And, unfortunately too,
the link to this note

<http://www.gellerlabs.com/34401A%20AC%20zero.htm>

doesn't work any more. Has anyone of you those pictures and could
provide them? Many thanks in advance.

Regards

Reinhard

Re: Low level AC measuremets with HP 34401

<7e4327c0-8491-4311-bb8b-8b09399517a5n@googlegroups.com>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85591&group=sci.electronics.design#85591

  copy link   Newsgroups: sci.electronics.design
X-Received: by 2002:a05:622a:2c9:: with SMTP id a9mr341884qtx.28.1640224957798;
Wed, 22 Dec 2021 18:02:37 -0800 (PST)
X-Received: by 2002:a25:2745:: with SMTP id n66mr567942ybn.120.1640224957493;
Wed, 22 Dec 2021 18:02:37 -0800 (PST)
Path: i2pn2.org!i2pn.org!weretis.net!feeder6.news.weretis.net!news.misty.com!border2.nntp.dca1.giganews.com!nntp.giganews.com!news-out.google.com!nntp.google.com!postnews.google.com!google-groups.googlegroups.com!not-for-mail
Newsgroups: sci.electronics.design
Date: Wed, 22 Dec 2021 18:02:37 -0800 (PST)
In-Reply-To: <j2i36iFhbirU1@mid.individual.net>
Injection-Info: google-groups.googlegroups.com; posting-host=94.145.255.13; posting-account=mW5JKwkAAAAMyuWOVeLp8yffyAkVx0g7
NNTP-Posting-Host: 94.145.255.13
References: <j2i36iFhbirU1@mid.individual.net>
User-Agent: G2/1.0
MIME-Version: 1.0
Message-ID: <7e4327c0-8491-4311-bb8b-8b09399517a5n@googlegroups.com>
Subject: Re: Low level AC measuremets with HP 34401
From: langw...@fonz.dk (Lasse Langwadt Christensen)
Injection-Date: Thu, 23 Dec 2021 02:02:37 +0000
Content-Type: text/plain; charset="UTF-8"
Lines: 19
 by: Lasse Langwadt Chris - Thu, 23 Dec 2021 02:02 UTC

torsdag den 23. december 2021 kl. 02.54.33 UTC+1 skrev Reinhard Zwirner:
> Hello,
>
> Due to the used RMS converter low level AC measurememts with this DMM
> are less precise to incorrect. There is a lab note as to this problem
> by Gellerlabs containing three pictures. I have this note which,
> unfortunately, lacks the mentioned pictures. And, unfortunately too,
> the link to this note
>
> <http://www.gellerlabs.com/34401A%20AC%20zero.htm>
>
> doesn't work any more. Has anyone of you those pictures and could
> provide them? Many thanks in advance.
>
> Regards
>
> Reinhard

google says, .. https://pearl-hifi.com/06_Lit_Archive/15_Mfrs_Publications/20_HP_Agilent/01_HP_3400B_20MHZ_True_RMS_Voltmeter/Geller%20Labs/34401A%20AC%20zero.pdf

Re: Low level AC measuremets with HP 34401

<23m7sg1bqkh1rravesc12gm6f54lbkvrej@4ax.com>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85592&group=sci.electronics.design#85592

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!weretis.net!feeder6.news.weretis.net!news.misty.com!border2.nntp.dca1.giganews.com!nntp.giganews.com!buffer2.nntp.dca1.giganews.com!nntp.supernews.com!news.supernews.com.POSTED!not-for-mail
NNTP-Posting-Date: Wed, 22 Dec 2021 20:03:23 -0600
From: jlar...@highlandsniptechnology.com
Newsgroups: sci.electronics.design
Subject: Re: Low level AC measuremets with HP 34401
Date: Wed, 22 Dec 2021 18:03:24 -0800
Message-ID: <23m7sg1bqkh1rravesc12gm6f54lbkvrej@4ax.com>
References: <j2i36iFhbirU1@mid.individual.net>
X-Newsreader: Forte Agent 3.1/32.783
MIME-Version: 1.0
Content-Type: text/plain; charset=us-ascii
Content-Transfer-Encoding: 7bit
Lines: 30
X-Trace: sv3-THmXfdVjeHL35tE0sgqyFH80POmEi4nQKJTwLaH/bMGZ4quW5+x26fvYXcm8U5+TakCXfHLQQ3B7Fwu!s+AI7Qrbao+tZ1gYp1hzpg6zuVvwMIP//45xk3OChqPcLl2uWUd1YBY8GKsTdU0qa/dR20f6UOwM!0EADxQ==
X-Complaints-To: www.supernews.com/docs/abuse.html
X-DMCA-Complaints-To: www.supernews.com/docs/dmca.html
X-Abuse-and-DMCA-Info: Please be sure to forward a copy of ALL headers
X-Abuse-and-DMCA-Info: Otherwise we will be unable to process your complaint properly
X-Postfilter: 1.3.40
X-Original-Bytes: 1858
 by: jlar...@highlandsniptechnology.com - Thu, 23 Dec 2021 02:03 UTC

On Thu, 23 Dec 2021 02:54:27 +0100, Reinhard Zwirner
<reinhard.zwirner@t-online.de> wrote:

>Hello,
>
>Due to the used RMS converter low level AC measurememts with this DMM
>are less precise to incorrect. There is a lab note as to this problem
>by Gellerlabs containing three pictures. I have this note which,
>unfortunately, lacks the mentioned pictures. And, unfortunately too,
>the link to this note
>
><http://www.gellerlabs.com/34401A%20AC%20zero.htm>
>
>doesn't work any more. Has anyone of you those pictures and could
>provide them? Many thanks in advance.
>
>Regards
>
>Reinhard

That DVM has a huge amount of noise kicked into the front end from the
VF display. They hid that by zeroing AC measurements much below 1 mV.

I like the Fluke better.

--

I yam what I yam - Popeye

Re: Low level AC measuremets with HP 34401

<87138eef-a707-4896-94fd-34d9533b8d4an@googlegroups.com>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85593&group=sci.electronics.design#85593

  copy link   Newsgroups: sci.electronics.design
X-Received: by 2002:a05:622a:1112:: with SMTP id e18mr364342qty.226.1640225851843;
Wed, 22 Dec 2021 18:17:31 -0800 (PST)
X-Received: by 2002:a25:707:: with SMTP id 7mr506313ybh.635.1640225851634;
Wed, 22 Dec 2021 18:17:31 -0800 (PST)
Path: i2pn2.org!i2pn.org!weretis.net!feeder6.news.weretis.net!news.misty.com!border2.nntp.dca1.giganews.com!nntp.giganews.com!news-out.google.com!nntp.google.com!postnews.google.com!google-groups.googlegroups.com!not-for-mail
Newsgroups: sci.electronics.design
Date: Wed, 22 Dec 2021 18:17:31 -0800 (PST)
In-Reply-To: <j2i36iFhbirU1@mid.individual.net>
Injection-Info: google-groups.googlegroups.com; posting-host=27.32.41.185; posting-account=B_tJMAoAAAAmar-1r2H3x4CMhbFEou3n
NNTP-Posting-Host: 27.32.41.185
References: <j2i36iFhbirU1@mid.individual.net>
User-Agent: G2/1.0
MIME-Version: 1.0
Message-ID: <87138eef-a707-4896-94fd-34d9533b8d4an@googlegroups.com>
Subject: Re: Low level AC measuremets with HP 34401
From: palliso...@gmail.com (Phil Allison)
Injection-Date: Thu, 23 Dec 2021 02:17:31 +0000
Content-Type: text/plain; charset="UTF-8"
Lines: 13
 by: Phil Allison - Thu, 23 Dec 2021 02:17 UTC

Reinhard Zwirner wrote:
==================
>
> Due to the used RMS converter low level AC measurememts with this DMM
> are less precise to incorrect.
>

** Amplify your signal prior to the DMM.
By 10x or 100X using say a wide band op-amp.

I use a 1000X pre amp for low level noise measurements.

...... Phil

Re: Low level AC measuremets with HP 34401

<20211222b@crcomp.net>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85595&group=sci.electronics.design#85595

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!eternal-september.org!reader02.eternal-september.org!.POSTED!not-for-mail
From: g...@crcomp.net (Don)
Newsgroups: sci.electronics.design
Subject: Re: Low level AC measuremets with HP 34401
Date: Thu, 23 Dec 2021 03:27:15 -0000 (UTC)
Organization: A noiseless patient Spider
Lines: 33
Message-ID: <20211222b@crcomp.net>
References: <j2i36iFhbirU1@mid.individual.net> <7e4327c0-8491-4311-bb8b-8b09399517a5n@googlegroups.com>
Content-Type: text/plain; charset=UTF-8
Content-Transfer-Encoding: 8bit
Injection-Date: Thu, 23 Dec 2021 03:27:15 -0000 (UTC)
Injection-Info: reader02.eternal-september.org; posting-host="fe92ed87da3b7dbfe98f75d1f45c2d19";
logging-data="12677"; mail-complaints-to="abuse@eternal-september.org"; posting-account="U2FsdGVkX19U/Hp47ipEnpSYGRxAe3w3"
Cancel-Lock: sha1:DQCcrOZjGFN8NGF64MHptKtEBqo=
 by: Don - Thu, 23 Dec 2021 03:27 UTC

Lasse Langwadt Christensen wrote:
> skrev Reinhard Zwirner:
>> Hello,
>>
>> Due to the used RMS converter low level AC measurememts with this DMM
>> are less precise to incorrect. There is a lab note as to this problem
>> by Gellerlabs containing three pictures. I have this note which,
>> unfortunately, lacks the mentioned pictures. And, unfortunately too,
>> the link to this note
>>
>> <http://www.gellerlabs.com/34401A%20AC%20zero.htm>
>>
>> doesn't work any more. Has anyone of you those pictures and could
>> provide them? Many thanks in advance.
>>
>> Regards
>>
>> Reinhard
>
>
> google says, .. https://pearl-hifi.com/06_Lit_Archive/15_Mfrs_Publications/20_HP_Agilent/01_HP_3400B_20MHZ_True_RMS_Voltmeter/Geller%20Labs/34401A%20AC%20zero.pdf

It's also available in HTML format at:

http://web.archive.org/web/20090322154522/http://www.gellerlabs.com/34401A%20AC%20zero.htm

Danke,

--
Don, KB7RPU, https://www.qsl.net/kb7rpu
There was a young lady named Bright Whose speed was far faster than light;
She set out one day In a relative way And returned on the previous night.

Re: Low level AC measuremets with HP 34401

<sq0vvm$6tg$1@dont-email.me>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85598&group=sci.electronics.design#85598

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!eternal-september.org!reader02.eternal-september.org!.POSTED!not-for-mail
From: blockedo...@foo.invalid (Don Y)
Newsgroups: sci.electronics.design
Subject: Re: Low level AC measuremets with HP 34401
Date: Wed, 22 Dec 2021 22:03:40 -0700
Organization: A noiseless patient Spider
Lines: 18
Message-ID: <sq0vvm$6tg$1@dont-email.me>
References: <j2i36iFhbirU1@mid.individual.net>
Mime-Version: 1.0
Content-Type: text/plain; charset=utf-8; format=flowed
Content-Transfer-Encoding: 7bit
Injection-Date: Thu, 23 Dec 2021 05:03:50 -0000 (UTC)
Injection-Info: reader02.eternal-september.org; posting-host="0ecb1be76d329cf3087ac7dbe9409496";
logging-data="7088"; mail-complaints-to="abuse@eternal-september.org"; posting-account="U2FsdGVkX1/G+Yk3RLjGjiXymZnJV8xF"
User-Agent: Mozilla/5.0 (Windows NT 6.1; WOW64; rv:52.0) Gecko/20100101
Thunderbird/52.1.1
Cancel-Lock: sha1:C+FNNqwy0frEVt4BCA4C7Ugc3Q4=
In-Reply-To: <j2i36iFhbirU1@mid.individual.net>
Content-Language: en-US
 by: Don Y - Thu, 23 Dec 2021 05:03 UTC

On 12/22/2021 6:54 PM, Reinhard Zwirner wrote:
> Hello,
>
> Due to the used RMS converter low level AC measurememts with this DMM
> are less precise to incorrect. There is a lab note as to this problem
> by Gellerlabs containing three pictures. I have this note which,
> unfortunately, lacks the mentioned pictures. And, unfortunately too,
> the link to this note
>
> <http://www.gellerlabs.com/34401A%20AC%20zero.htm>
>
> doesn't work any more. Has anyone of you those pictures and could
> provide them? Many thanks in advance.

Try:

<https://web.archive.org/web/20090106175105/http://www.gellerlabs.com/34401A%20AC%20zero.htm>

Re: Low level AC measuremets with HP 34401

<0a7ec75a-c8c6-4c1b-8763-2420a2401751n@googlegroups.com>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85600&group=sci.electronics.design#85600

  copy link   Newsgroups: sci.electronics.design
X-Received: by 2002:ac8:5f47:: with SMTP id y7mr684961qta.342.1640240094061;
Wed, 22 Dec 2021 22:14:54 -0800 (PST)
X-Received: by 2002:a05:6902:1145:: with SMTP id p5mr1262883ybu.327.1640240093824;
Wed, 22 Dec 2021 22:14:53 -0800 (PST)
Path: i2pn2.org!i2pn.org!weretis.net!feeder6.news.weretis.net!news.misty.com!border2.nntp.dca1.giganews.com!nntp.giganews.com!news-out.google.com!nntp.google.com!postnews.google.com!google-groups.googlegroups.com!not-for-mail
Newsgroups: sci.electronics.design
Date: Wed, 22 Dec 2021 22:14:53 -0800 (PST)
In-Reply-To: <87138eef-a707-4896-94fd-34d9533b8d4an@googlegroups.com>
Injection-Info: google-groups.googlegroups.com; posting-host=27.32.41.185; posting-account=B_tJMAoAAAAmar-1r2H3x4CMhbFEou3n
NNTP-Posting-Host: 27.32.41.185
References: <j2i36iFhbirU1@mid.individual.net> <87138eef-a707-4896-94fd-34d9533b8d4an@googlegroups.com>
User-Agent: G2/1.0
MIME-Version: 1.0
Message-ID: <0a7ec75a-c8c6-4c1b-8763-2420a2401751n@googlegroups.com>
Subject: Re: Low level AC measuremets with HP 34401
From: palliso...@gmail.com (Phil Allison)
Injection-Date: Thu, 23 Dec 2021 06:14:54 +0000
Content-Type: text/plain; charset="UTF-8"
Lines: 26
 by: Phil Allison - Thu, 23 Dec 2021 06:14 UTC

Phil Allison wrote:
==============
> Reinhard Zwirner wrote:
> ==================
> >
> > Due to the used RMS converter low level AC measurememts with this DMM
> > are less precise to incorrect.
> >
> ** Amplify your signal prior to the DMM.
> By 10x or 100X using say a wide band op-amp.
>
> I use a 1000X pre amp for low level noise measurements.
>

** FYI:

I have a 4.5. digit ( 20,000 count ) DMM that employs an AD536 rms-DC converter - cousin of the AD637 in the HP.
The converter IC has bandwidth inversely proportional to the input voltage, so 10% of range is a good place to stop thinking it very accurate.
Using one with a 4.5 digit meter is pushing it, so 6.5 is off the map.

Suspect the HP already has 10X and maybe 100X gain stages internally, prior to the converter.
In any case, an external pre-amp is the go.

...... Phil

Thank you! (was: Low level AC measuremets with HP 34401)

<j2j3pgFn4j0U1@mid.individual.net>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85616&group=sci.electronics.design#85616

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!paganini.bofh.team!news.dns-netz.com!news.freedyn.de!newsreader4.netcologne.de!news.netcologne.de!fu-berlin.de!uni-berlin.de!individual.net!not-for-mail
From: reinhard...@t-online.de (Reinhard Zwirner)
Newsgroups: sci.electronics.design
Subject: Thank you! (was: Low level AC measuremets with HP 34401)
Date: Thu, 23 Dec 2021 12:10:41 +0100
Lines: 7
Message-ID: <j2j3pgFn4j0U1@mid.individual.net>
References: <j2i36iFhbirU1@mid.individual.net>
Mime-Version: 1.0
Content-Type: text/plain; charset=UTF-8
Content-Transfer-Encoding: 7bit
X-Trace: individual.net r42gnOQ/RJdBbm+Y3MFsmwGzda6rCY1gzoKbYp4+IfEI8=
Cancel-Lock: sha1:60sYQLwwA+0uvoSoCg/NhZSXb8I=
User-Agent: Mozilla/5.0 (Windows NT 10.0; Win64; x64; rv:68.0) Gecko/20100101
Firefox/68.0 SeaMonkey/2.53.10.1
In-Reply-To: <j2i36iFhbirU1@mid.individual.net>
 by: Reinhard Zwirner - Thu, 23 Dec 2021 11:10 UTC

Hello,

Many thanks to all helpers :-)!

Merry Chrismas and a Happy New Year

Reinhard

Re: Low level AC measuremets with HP 34401

<tscheppe.i0b8oov53dpl@nntp.aioe.org>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85657&group=sci.electronics.design#85657

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!aioe.org!izCICPyInj/8aZGhGyo7Pg.user.46.165.242.75.POSTED!not-for-mail
From: klausk...@hotmail.com (Klaus Kragelund)
Newsgroups: sci.electronics.design
Subject: Re: Low level AC measuremets with HP 34401
Date: Thu, 23 Dec 2021 20:50:08 +0100
Organization: Aioe.org NNTP Server
Message-ID: <tscheppe.i0b8oov53dpl@nntp.aioe.org>
References: <j2i36iFhbirU1@mid.individual.net> <23m7sg1bqkh1rravesc12gm6f54lbkvrej@4ax.com>
Mime-Version: 1.0
Content-Type: text/plain; charset=utf-8; format=flowed
Content-Transfer-Encoding: 8bit
Injection-Info: gioia.aioe.org; logging-data="58895"; posting-host="izCICPyInj/8aZGhGyo7Pg.user.gioia.aioe.org"; mail-complaints-to="abuse@aioe.org";
User-Agent: Newsgroups for Android by Tscheppe
X-Notice: Filtered by postfilter v. 0.9.2
Content-Language: en_DK
 by: Klaus Kragelund - Thu, 23 Dec 2021 19:50 UTC

23.12.21 03:03, jlarkin@highlandsniptechnology.com wrote:
>On Thu, 23 Dec 2021 02:54:27 +0100, Reinhard Zwirner
><reinhard.zwirner@t-online.de> wrote:
>
>>Hello,
>>
>>Due to the used RMS converter low level AC measurememts with this DMM
>>are less precise to incorrect. There is a lab note as to this problem
>>by Gellerlabs containing three pictures. I have this note which,
>>unfortunately, lacks the mentioned pictures. And, unfortunately too,
>>the link to this note
>>
>><http://www.gellerlabs.com/34401A%20AC%20zero.htm>
>>
>>doesn't work any more. Has anyone of you those pictures and could
>>provide them? Many thanks in advance.
>>
>>Regards
>>
>>Reinhard
>
>That DVM has a huge amount of noise kicked into the front end from the
>VF display. They hid that by zeroing AC measurements much below 1 mV.
>
>I like the Fluke better.
>

The 34401A was my first bench DMM. Bought it in 1998.

I have never calibrated it, could not afford it

This year I went the consultant route, so bought 3 Siglent DMMs. Just for fun, I connected the same references voltage to the 3 Siglent and the 34401A
They measured the same except for minor last digit difference

Not willing to spend on calibrations, the next best thing is to compare to other instruments. Quite unlikely that they would drift in the same direction ?

By the way, I paid a lot for the 34401A, now I can get 4 Siglents for the same price

--
Klaus

Re: Low level AC measuremets with HP 34401

<sq2m53$lgf$1@reader1.panix.com>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85664&group=sci.electronics.design#85664

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!weretis.net!feeder6.news.weretis.net!panix!.POSTED.panix2.panix.com!not-for-mail
From: prese...@MUNGEpanix.com (Cydrome Leader)
Newsgroups: sci.electronics.design
Subject: Re: Low level AC measuremets with HP 34401
Date: Thu, 23 Dec 2021 20:28:19 -0000 (UTC)
Organization: PANIX Public Access Internet and UNIX, NYC
Message-ID: <sq2m53$lgf$1@reader1.panix.com>
References: <j2i36iFhbirU1@mid.individual.net> <23m7sg1bqkh1rravesc12gm6f54lbkvrej@4ax.com> <tscheppe.i0b8oov53dpl@nntp.aioe.org>
Injection-Date: Thu, 23 Dec 2021 20:28:19 -0000 (UTC)
Injection-Info: reader1.panix.com; posting-host="panix2.panix.com:166.84.1.2";
logging-data="22031"; mail-complaints-to="abuse@panix.com"
User-Agent: tin/2.6.0-20210823 ("Coleburn") (NetBSD/9.2 (amd64))
 by: Cydrome Leader - Thu, 23 Dec 2021 20:28 UTC

Klaus Kragelund <klauskvik@hotmail.com> wrote:
> 23.12.21 03:03, jlarkin@highlandsniptechnology.com wrote:
>>On Thu, 23 Dec 2021 02:54:27 +0100, Reinhard Zwirner
>><reinhard.zwirner@t-online.de> wrote:
>>
>>>Hello,
>>>
>>>Due to the used RMS converter low level AC measurememts with this DMM
>>>are less precise to incorrect. There is a lab note as to this problem
>>>by Gellerlabs containing three pictures. I have this note which,
>>>unfortunately, lacks the mentioned pictures. And, unfortunately too,
>>>the link to this note
>>>
>>><http://www.gellerlabs.com/34401A%20AC%20zero.htm>
>>>
>>>doesn't work any more. Has anyone of you those pictures and could
>>>provide them? Many thanks in advance.
>>>
>>>Regards
>>>
>>>Reinhard
>>
>>That DVM has a huge amount of noise kicked into the front end from the
>>VF display. They hid that by zeroing AC measurements much below 1 mV.
>>
>>I like the Fluke better.
>>
>
> The 34401A was my first bench DMM. Bought it in 1998.
>
> I have never calibrated it, could not afford it
>
> This year I went the consultant route, so bought 3 Siglent DMMs. Just for fun, I connected the same references voltage to the 3 Siglent and the 34401A
> They measured the same except for minor last digit difference
>
> Not willing to spend on calibrations, the next best thing is to compare to other instruments. Quite unlikely that they would drift in the same direction ?
>
> By the way, I paid a lot for the 34401A, now I can get 4 Siglents for the same price
>
>
> --
> Klaus

I like the 34401, even with some of the faults. As you've seen they're
basically accurate forever. Not sure how the new keysigh stuff compares
for long term reliability.

Re: Low level AC measuremets with HP 34401

<tscheppe.c2z4telc5n1z@nntp.aioe.org>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85666&group=sci.electronics.design#85666

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!aioe.org!izCICPyInj/8aZGhGyo7Pg.user.46.165.242.75.POSTED!not-for-mail
From: klausk...@hotmail.com (Klaus Kragelund)
Newsgroups: sci.electronics.design
Subject: Re: Low level AC measuremets with HP 34401
Date: Thu, 23 Dec 2021 21:38:25 +0100
Organization: Aioe.org NNTP Server
Message-ID: <tscheppe.c2z4telc5n1z@nntp.aioe.org>
References: <j2i36iFhbirU1@mid.individual.net> <23m7sg1bqkh1rravesc12gm6f54lbkvrej@4ax.com> <tscheppe.i0b8oov53dpl@nntp.aioe.org> <sq2m53$lgf$1@reader1.panix.com>
Mime-Version: 1.0
Content-Type: text/plain; charset=utf-8; format=flowed
Content-Transfer-Encoding: 8bit
Injection-Info: gioia.aioe.org; logging-data="34409"; posting-host="izCICPyInj/8aZGhGyo7Pg.user.gioia.aioe.org"; mail-complaints-to="abuse@aioe.org";
User-Agent: Newsgroups for Android by Tscheppe
Content-Language: en_DK
X-Notice: Filtered by postfilter v. 0.9.2
 by: Klaus Kragelund - Thu, 23 Dec 2021 20:38 UTC

23.12.21 21:28, Cydrome Leader wrote:
>Klaus Kragelund <klauskvik@hotmail.com> wrote:
>> 23.12.21 03:03, jlarkin@highlandsniptechnology.com wrote:
>>>On Thu, 23 Dec 2021 02:54:27 +0100, Reinhard Zwirner
>>><reinhard.zwirner@t-online.de> wrote:
>>>
>>>>Hello,
>>>>
>>>>Due to the used RMS converter low level AC measurememts with this DMM
>>>>are less precise to incorrect. There is a lab note as to this problem
>>>>by Gellerlabs containing three pictures. I have this note which,
>>>>unfortunately, lacks the mentioned pictures. And, unfortunately too,
>>>>the link to this note
>>>>
>>>><http://www.gellerlabs.com/34401A%20AC%20zero.htm>
>>>>
>>>>doesn't work any more. Has anyone of you those pictures and could
>>>>provide them? Many thanks in advance.
>>>>
>>>>Regards
>>>>
>>>>Reinhard
>>>
>>>That DVM has a huge amount of noise kicked into the front end from the
>>>VF display. They hid that by zeroing AC measurements much below 1 mV.
>>>
>>>I like the Fluke better.
>>>
>>
>> The 34401A was my first bench DMM. Bought it in 1998.
>>
>> I have never calibrated it, could not afford it
>>
>> This year I went the consultant route, so bought 3 Siglent DMMs. Just for fun, I connected the same references voltage to the 3 Siglent and the 34401A
>> They measured the same except for minor last digit difference
>>
>> Not willing to spend on calibrations, the next best thing is to compare to other instruments. Quite unlikely that they would drift in the same direction ?
>>
>> By the way, I paid a lot for the 34401A, now I can get 4 Siglents for the same price
>>
>>
>> --
>> Klaus
>
>I like the 34401, even with some of the faults. As you've seen they're
>basically accurate forever.

At the place I worked before they used it per default. I asked the instrument responsible and he said when they were sent for calibration they were never out of spec. Not even a single instrument. So they prolonged the calibration interval to 2 years

--
Klaus

Re: Low level AC measuremets with HP 34401

<1pv9sgdtgnugtvemd06m68cdga6t7mh3pa@4ax.com>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85671&group=sci.electronics.design#85671

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!weretis.net!feeder6.news.weretis.net!news.misty.com!border2.nntp.dca1.giganews.com!nntp.giganews.com!buffer2.nntp.dca1.giganews.com!buffer1.nntp.dca1.giganews.com!nntp.supernews.com!news.supernews.com.POSTED!not-for-mail
NNTP-Posting-Date: Thu, 23 Dec 2021 17:01:27 -0600
From: jlar...@highland_atwork_technology.com (John Larkin)
Newsgroups: sci.electronics.design
Subject: Re: Low level AC measuremets with HP 34401
Date: Thu, 23 Dec 2021 15:01:27 -0800
Organization: Highland Tech
Reply-To: xx@yy.com
Message-ID: <1pv9sgdtgnugtvemd06m68cdga6t7mh3pa@4ax.com>
References: <j2i36iFhbirU1@mid.individual.net> <23m7sg1bqkh1rravesc12gm6f54lbkvrej@4ax.com> <tscheppe.i0b8oov53dpl@nntp.aioe.org> <sq2m53$lgf$1@reader1.panix.com>
X-Newsreader: Forte Agent 3.1/32.783
MIME-Version: 1.0
Content-Type: text/plain; charset=us-ascii
Content-Transfer-Encoding: 7bit
Lines: 60
X-Trace: sv3-4OZ6FRpDJ0UpsJHwCsNiboUoymtDMc0DeEzrIzj2EV/otVE3JnE0NaSV+v76Jj++RoWMsBWJHzDcwWG!pj+eejOVRBi6ILVc/kIzQXCIzrR5wBxnaHYsw6GCj+z3Wx5f1TToFOuMP93DRp/J1p5p4hr2ta/w!FCdfbQ==
X-Complaints-To: www.supernews.com/docs/abuse.html
X-DMCA-Complaints-To: www.supernews.com/docs/dmca.html
X-Abuse-and-DMCA-Info: Please be sure to forward a copy of ALL headers
X-Abuse-and-DMCA-Info: Otherwise we will be unable to process your complaint properly
X-Postfilter: 1.3.40
X-Original-Bytes: 3419
 by: John Larkin - Thu, 23 Dec 2021 23:01 UTC

On Thu, 23 Dec 2021 20:28:19 -0000 (UTC), Cydrome Leader
<presence@MUNGEpanix.com> wrote:

>Klaus Kragelund <klauskvik@hotmail.com> wrote:
>> 23.12.21 03:03, jlarkin@highlandsniptechnology.com wrote:
>>>On Thu, 23 Dec 2021 02:54:27 +0100, Reinhard Zwirner
>>><reinhard.zwirner@t-online.de> wrote:
>>>
>>>>Hello,
>>>>
>>>>Due to the used RMS converter low level AC measurememts with this DMM
>>>>are less precise to incorrect. There is a lab note as to this problem
>>>>by Gellerlabs containing three pictures. I have this note which,
>>>>unfortunately, lacks the mentioned pictures. And, unfortunately too,
>>>>the link to this note
>>>>
>>>><http://www.gellerlabs.com/34401A%20AC%20zero.htm>
>>>>
>>>>doesn't work any more. Has anyone of you those pictures and could
>>>>provide them? Many thanks in advance.
>>>>
>>>>Regards
>>>>
>>>>Reinhard
>>>
>>>That DVM has a huge amount of noise kicked into the front end from the
>>>VF display. They hid that by zeroing AC measurements much below 1 mV.
>>>
>>>I like the Fluke better.
>>>
>>
>> The 34401A was my first bench DMM. Bought it in 1998.
>>
>> I have never calibrated it, could not afford it
>>
>> This year I went the consultant route, so bought 3 Siglent DMMs. Just for fun, I connected the same references voltage to the 3 Siglent and the 34401A
>> They measured the same except for minor last digit difference
>>
>> Not willing to spend on calibrations, the next best thing is to compare to other instruments. Quite unlikely that they would drift in the same direction ?
>>
>> By the way, I paid a lot for the 34401A, now I can get 4 Siglents for the same price
>>
>>
>> --
>> Klaus
>
>I like the 34401, even with some of the faults. As you've seen they're
>basically accurate forever. Not sure how the new keysigh stuff compares
>for long term reliability.

We use mostly Keithleys in our test stands. They're OK.

I really like my Fluke 8845A. It's the winner when you want to measure
a few microvolts.

--

If a man will begin with certainties, he shall end with doubts,
but if he will be content to begin with doubts he shall end in certainties.
Francis Bacon

Re: Low level AC measuremets with HP 34401

<sq372q$653$1@reader1.panix.com>

  copy mid

https://www.novabbs.com/tech/article-flat.php?id=85678&group=sci.electronics.design#85678

  copy link   Newsgroups: sci.electronics.design
Path: i2pn2.org!i2pn.org!weretis.net!feeder6.news.weretis.net!panix!.POSTED.panix2.panix.com!not-for-mail
From: prese...@MUNGEpanix.com (Cydrome Leader)
Newsgroups: sci.electronics.design
Subject: Re: Low level AC measuremets with HP 34401
Date: Fri, 24 Dec 2021 01:17:14 -0000 (UTC)
Organization: PANIX Public Access Internet and UNIX, NYC
Message-ID: <sq372q$653$1@reader1.panix.com>
References: <j2i36iFhbirU1@mid.individual.net> <23m7sg1bqkh1rravesc12gm6f54lbkvrej@4ax.com> <tscheppe.i0b8oov53dpl@nntp.aioe.org> <sq2m53$lgf$1@reader1.panix.com> <tscheppe.c2z4telc5n1z@nntp.aioe.org>
Injection-Date: Fri, 24 Dec 2021 01:17:14 -0000 (UTC)
Injection-Info: reader1.panix.com; posting-host="panix2.panix.com:166.84.1.2";
logging-data="6307"; mail-complaints-to="abuse@panix.com"
User-Agent: tin/2.6.0-20210823 ("Coleburn") (NetBSD/9.2 (amd64))
 by: Cydrome Leader - Fri, 24 Dec 2021 01:17 UTC

Klaus Kragelund <klauskvik@hotmail.com> wrote:
> 23.12.21 21:28, Cydrome Leader wrote:
>>Klaus Kragelund <klauskvik@hotmail.com> wrote:
>>> 23.12.21 03:03, jlarkin@highlandsniptechnology.com wrote:
>>>>On Thu, 23 Dec 2021 02:54:27 +0100, Reinhard Zwirner
>>>><reinhard.zwirner@t-online.de> wrote:
>>>>
>>>>>Hello,
>>>>>
>>>>>Due to the used RMS converter low level AC measurememts with this DMM
>>>>>are less precise to incorrect. There is a lab note as to this problem
>>>>>by Gellerlabs containing three pictures. I have this note which,
>>>>>unfortunately, lacks the mentioned pictures. And, unfortunately too,
>>>>>the link to this note
>>>>>
>>>>><http://www.gellerlabs.com/34401A%20AC%20zero.htm>
>>>>>
>>>>>doesn't work any more. Has anyone of you those pictures and could
>>>>>provide them? Many thanks in advance.
>>>>>
>>>>>Regards
>>>>>
>>>>>Reinhard
>>>>
>>>>That DVM has a huge amount of noise kicked into the front end from the
>>>>VF display. They hid that by zeroing AC measurements much below 1 mV.
>>>>
>>>>I like the Fluke better.
>>>>
>>>
>>> The 34401A was my first bench DMM. Bought it in 1998.
>>>
>>> I have never calibrated it, could not afford it
>>>
>>> This year I went the consultant route, so bought 3 Siglent DMMs. Just for fun, I connected the same references voltage to the 3 Siglent and the 34401A
>>> They measured the same except for minor last digit difference
>>>
>>> Not willing to spend on calibrations, the next best thing is to compare to other instruments. Quite unlikely that they would drift in the same direction ?
>>>
>>> By the way, I paid a lot for the 34401A, now I can get 4 Siglents for the same price
>>>
>>>
>>> --
>>> Klaus
>>
>>I like the 34401, even with some of the faults. As you've seen they're
>>basically accurate forever.
>
> At the place I worked before they used it per default. I asked the
> instrument responsible and he said when they were sent for calibration
> they were never out of spec. Not even a single instrument. So they
> prolonged the calibration interval to 2 years
>
>
> --
> Klaus

Calibration is mostly just a racket, although I completely agree you need
to check instruments to make sure they're not broken. I have the hardest
time with stuff like capacitance meters.

As for calibration of things like rulers, I confirmed that yes, I can hand
draw a 1 foot ruler that's only 11 inches long on a paper towel and sent
it it for calibration at a metrology lab, and that with the calibration
chart it would be be considered accurate and for a few bucks more I could
claim NIST traceable for my fine instrument. Would it be useful? No, as
long as all the errors are documented, it's better than a Starret scale
with no paperwork, at least in the eyes of retarded places ike NASA.